Pressure Control Ventilation Monitoring

Key Points

  • In pressure-control ventilation, inspiratory pressure is fixed and delivered tidal volume varies with lung mechanics.
  • Continuous trend monitoring of delivered volumes is essential to confirm effective and safe ventilation.
  • Changes in compliance or resistance can rapidly alter volume delivery without a setting change.

Equipment

  • Ventilator running AC pressure-control mode
  • Bedside display/trending for tidal volume, inspiratory flow, and I:E ratio
  • Current order set including RR, pressure control level, inspiratory time, PEEP, and FiO2
  • Structured handoff documentation for trend comparison over time

Procedure Steps

  1. Confirm the patient is on pressure-control mode and verify ordered set parameters.
  2. Verify directly set parameters: respiratory rate, pressure control, inspiratory time, PEEP, and FiO2.
  3. Identify key observed variables that are not directly set: tidal volume, inspiratory flow, and I:E.
  4. Establish baseline observed values immediately after settings are confirmed.
  5. Trend delivered tidal volume over time to detect improvement or deterioration in lung mechanics.
  6. Correlate abrupt tidal-volume changes with possible compliance/resistance changes and reassess patient status.
  7. Notify respiratory therapy/provider promptly when delivered volume is no longer in a safe/effective range.
  8. Reassess after each setting change and document mode, set values, observed trends, and response.
  9. Continue repeated trend checks at routine intervals and during clinical deterioration.

Common Errors

  • Assuming fixed pressure guarantees stable ventilation missed low- or high-volume risk.
  • Failing to trend tidal volume over time delayed recognition of worsening compliance/resistance.
  • Ignoring observed I:E and flow changes incomplete assessment of ventilator-patient interaction.
  • Delayed escalation after unsafe volume trend preventable hypoventilation or lung injury.